Automated Cleveland Open Cup Flash Point Testeraco-8 / aco-8e / aco-8as / aco-8eas

Automated Cleveland Open Cup Flash Point Testeraco-8 / aco-8e / aco-8as / aco-8eas

Since 1966, Tanaka has been delivering automated flash point testers for higher productivity at petroleum laboratories. The theme of the 8th generation model series is “ Ergonomic, Versatile, Robust and Safe “

●Ergonomic: Color-coordinated display and switches for high visibility / intuitive operation, and universal design
●Versatile: USB port for USB flash drive or keyboard, password protection and enhanced data storage
●Robust: Premium design and high-quality finish suited for many years of reliable use
●Safe: Enhanced safety with fire containment system (Default for apm, atg & aco, optional for abl)

Further improved Swing Arm Assembly for easy setting and
cleaning :
After inserting the test cup in the instrument, just lower the swing arm assembly onto the test cup to set all the necessary components for the test, including the ignition source, temperature sensor, flash detector and stirrer (in case of apm, atg and abl series). After the test, lift the arm to the complete upright position for easy cleaning.

aco-8 is an automated Cleveland Open Cup (COC) flash & fire point tester.
Flash point as well as fire point can be determined automatically. The fire point is detected when the double ionization rings detect 5 sec of continuous burning.
In addition to the existing gas ignition models (aco-8/8as), 2 new electric ignition models (aco-8e/8eas) have been launched. An electric igniter coil replaces the gas flame which allows the user to remove combustible gas from the lab and operate in a safer environment.
Fire containment by a metallic lid covers the test cup automatically when a sustained fire is detected. Model aco-8as and aco-8eas are for testing bituminous materials.

Specifications

aco-8aco-8asaco-8eaco-8eas
Test MethodsASTM D92, ISO 2592, IP 36,
JIS K 2265-4, AASHTO T48
JIS K 2265-4
Related Test MethodsASTM D92, ISO 2592, IP 36, AASHTO T48
Measuring RangeAmbient to 400 °C
Temperature SensorPT-100 in stainless steel sheath
Flash DetectorDouble ionization rings
Ignition SourceGas ignition with automatic lightingElectric ignition coil
I/O PortRC-232C x 1, USB x 1
Size (W x D x H)230 mm x 470 mm x 400 mm
Weight17 kg
Power ConsumptionAC 100 V to 120 V 50 / 60 Hz 1500 VA(max)
AC 220 V to 240 V 50 / 60 Hz 1300 VA(max)
Catalog