Automated Tag Closed Cup Flash Point Testeratg-8wfc / atg-8afc / atg-8lfc

Automated Tag Closed Cup Flash Point Testeratg-8wfc / atg-8afc / atg-8lfc

Since 1966, Tanaka has been delivering automated flash point testers for higher productivity at petroleum laboratories. The theme of the 8th generation model series is “ Ergonomic, Versatile, Robust and Safe “

●Ergonomic: Color-coordinated display and switches for high visibility / intuitive operation, and universal design
●Versatile: USB port for USB flash drive or keyboard, password protection and enhanced data storage
●Robust: Premium design and high-quality finish suited for many years of reliable use
●Safe: Enhanced safety with fire containment system (Default for apm, atg & aco, optional for abl)

Further improved Swing Arm Assembly for easy setting and
cleaning :
After inserting the test cup in the instrument, just lower the swing arm assembly onto the test cup to set all the necessary components for the test, including the ignition source, temperature sensor, flash detector and stirrer (in case of apm, atg and abl series). After the test, lift the arm to the complete upright position for easy cleaning.

atg-8 is an automated Tag Closed Cup flash point tester.
Model atg-8wfc is with a conventional water bath. Model atg-8afc and atg-8lfc are available for low-temperature range. The cooling/heating for these models is by Peltier modules (Model at-8lfc requires an optional chiller).
Furthermore, it is standard equipped with a fire containment system. It can extinguish fire effectively by flowing quenching gas (N2 or CO2) from the external connection port.

Specifications Catalog

Test MethodsASTM D56, IP 304, JIS K2265-1
Measuring RangeAmbient to +95 °C for atg-8wfc
+5 °C to +75 °C for atg-8afc
-20 °C to +95 °C for atg-8lfc (with an optional chiller)
Temperature SensorPT-100 in stainless steel sheath
Flash DetectorCRC thermocouple
Ignition SourceGas flame or electric coil
I/O PortRC-232C x 1, USB x 1
Size (W x D x H) 230 mm x 470 mm x 390 mm
Weight19 kg
Power Consumptionatg-8wfc: AC 100 V to 120 V 50 / 60 Hz 500 VA (max)
       AC 220 V to 240 V 50 / 60 Hz 400 VA (max)
atg-8afc / atg-8lfc: AC 100 V to 240 V 50 / 60 Hz 300 VA (max)